Manufacturer
|
Amphenol |
Series
|
T8041 |
Measurement Range
|
0-2000 PPM factory calibrated |
Duct Air Velocity
|
0 to 1500 ft/min (0 to 450 meter/min) |
Temperature Dependance
|
0.2% of full scale per ºC |
Stability
|
<2% of full scale over the life of the sensor (10 years typical) |
Accuracy
|
±40 ppm +3% of reading @ 22ºC (72ºF) when compared with a factory certified reference |
Non-Linearity
|
<1% of full scale @ 22ºC (72ºF) |
Pressure Dependence
|
0.13% of reading per mm of mercury |
Response Time
|
3 minutes typical for a 90% step change at low duct speeds |
Sampling Rate
|
Every two seconds, 25% duty cycle |
Operating Conditions
|
Temperature: 0ºC to 50ºC (32ºF to 122ºF) Humidity: 0 to 95% relative humidity, noncondensing |
Storage Conditions
|
-20ºC to 70ºC (-40ºF to 158ºF) |
Note
|
The Telaire product line offers patented ABC Logic™ software for self-correction of drift to better than ±20 ppm per year. The system is virtually free of maintenance and typically has than 10 years. |