Manufacturer
|
Amphenol |
Series
|
T8100-D-R |
Sensing Method
|
Gold-plated optics, Non-dispersive infrared (NDIR) absorbsion, Patented ABC Logic self calibration algorithm |
Measurement Range
|
0 to 5000 ppm** |
Dimensions
|
Overall 60 mm X 35 mm X 15 mm approx |
Accuracy
|
400-5000 ppm ± 75 ppm or 10% of reading, whichever is greater* |
Temperature Dependance
|
0.2% FS per °C |
Stability
|
<2% of FS over life of sensor (15 years) |
Pressure Dependence
|
0.135% of reading per mm Hg |
Calibration Interval
|
Not Required |
Response Time
|
7 minutes (approx) |
Signal Update
|
Every 4 seconds |
Warm Up TIme
|
< 2 minutes (operational) 10 minutes (maximum accuracy) |
Operating Conditions
|
32°F to 122°F (0°C to 50°C) 0 to 95% RH, non-condensing |
Storage Conditions
|
-40°F to 158°F (-40°C to 70°C) |
Outputs
|
I2C Digital, see T6700 Series Application Guide for detail |
Protection
|
IP65 |
Power Supply Requirements
|
4.5-5.5 VDC Peak 200mA (155mA typical) Average 25mA (20mA typical) |
Warranty Terms
|
12 Months |
Compliance
|
REACH, ROSH and Prop65 compliance Flammability rating of enclosure UL94 5VA |
Note
|
* Tolerance based on span gas of ±2% ** Subjecting sensors to environments less than 400ppm for more than 15 minutes may affectaccuracy due to ABC Logic™ algorithm. |